pHメーターの酸性・アルカリ度センサー電極の選択
1. シェル材質の選択:
pH電極の筐体は一般にPCプラスチック(ポリカーボネート)筐体とガラス筐体の2種類が採用されています。 PC ケースは衝突や衝撃に強いですが、温度にも適しています。<80oC. It is also susceptible to corrosion in high alkaline solutions and some media. The glass shell is suitable for temperatures ranging from 0-150oC and is generally not corroded except by hydrofluoric acid solution, but it is prone to collision damage.
2. 液体界面の選択:
The liquid interface is a connecting component that communicates the external reference solution and the measured solution, requiring stable and reproducible potential. General liquid interface materials include fiber materials, ceramic cores, glass grinding ports, and other media. Fiber materials are generally used in plastic shell electrodes, and their solution leakage rate is fast and not easily blocked; Ceramic cores have good corrosion resistance, and the stability and reproducibility of the liquid interface potential are also good. They can be used in high-temperature media and are widely used as liquid interface materials; The glass grinding interface has a large contact area and leakage rate with the solution, making it suitable for the determination of weak ionic strength, high viscosity, turbid liquids or colloidal solutions.
3. pH 感受性ガラスフィルムの選択:
PH感応性ガラスフィルムは、水素機能を持ったリチウムガラスを溶融・吹き飛ばして作られており、一般に球状の形状をしています。ほとんどの電極には従来の高感度ガラスフィルムが使用されていますが、高温媒体用、高温強酸媒体用、高温蒸気(130℃)用、低抵抗の高感度ガラスフィルム(純水測定用)などもあり、状況に応じて選択する必要があります。
4. 参照電極の選択:
Ag/AgCl 参照電極が一般的に使用されますが、一般的な Ag/AgCl 参照電極は高温で溶解しやすく、電位が不安定です。一方、キャピラリー構造の Ag/AgCl 参照電極は非常に安定した参照電位を持ち、高温および長期の連続試験条件下での使用に適しています。
5. 電極ソケットの選択:
pH電極のソケットはBNC型(別名Q9型)が一般的ですが、その他にも対応する機器とのマッチングに応じて様々な形状があります。電極モデルの最後の文字はソケットのタイプを示します。